■B1517A (X4) - Four HRSMU - Up to 100V, 100mA force, 1fA current resolution
■B1520A - MFCMU - 1kHz to 5MHz, up to 100V DC bias with SMU GET QUg-12
■B1517A (X4) - Four HRSMU - Up to 100V, 100mA force, 1fA current resolution
■B1530A - WGFMU - Min 100ns pulse width, 10V peak-to-peak output, 5ns current or voltage measurement sampling speedET QUOTE
Product Overview
The Keysight B1500A (Agilent) Semiconductor Device Analyzer is a modular instrument with a ten-slot configuration that supports both IV and CV measurements and also fast high-voltage pulsing. Its familiar, Microsoft ® Windows ® user interface supports Agilent’s EasyEXPERT software, which provides a new, more intuitive task-oriented approach to device characterization. Because of its extremely low-current, low-voltage, and integrated capacitance measurement capabilities, the Keysight B1500A (Agilent) can be used for a wide range of semiconductor device characterization needs (IC-CAP supports the B1500A). It is also an excellent solution for non-volatile memory cell characterization and high-speed device characterization (including advanced NBTI measurement).
Features and benefits of:
●Superior IV measurement performance: 0.1 fA / 0.5 µV measurement resolution
●Measurement features include single and multi-channel sweep, time sampling, list sweep, quasi-static CV (using the SMUs), direct control and arbitrary linear waveform generation (ALWG) GUI for the HV-SPGUs.
●Optional, integrated capacitance module supports CV measurements up to 5 MHz
●Optional positioner-based CV-IV switching solutions available with 0.5 µV voltage measurement resolution and 10 fA, 1 fA or 0.1 fA current measurement resolution capability
●Easy test automation with built-in semiautomatic wafer prober drivers and test sequencing without programming via the Quick Test mode
●Optional high-voltage semiconductor pulse generator unit (HV-SPGU) available with 10 ns programmable pulse widths and +/- 40 V (80 V peak-to-peak) output.
●Optional waveform generator/fast measurement unit (WGFMU) available with ALWG and fast current or voltage measurement capabilities.
●10 ns pulsed IV solution is available for characterizing high-k gate dielectric and SOI (silicon-on-insulator) transistors.
●A Classic Test mode is available to provide the look, feel, and terminology of the 4155/4156 interface while enhancing user interaction by taking full advantage of Microsoft ® Windows ® GUI features.
Key specifications of:
●General Features
●PC-based instrument with Windows ® XP Professional OS
●Single-box solution for current-voltage (IV), capacitance-voltage (CV), pulse generation, fast IV, and time-domain measurement.
●Ten module slots for source monitor units (SMUs) and other module types (MFCMU, HV-SPGU and WGFMU) *
●Offline data analysis and application test development via Desktop EasyEXPERT software
●Measurement Capabilities
■Supports current-voltage (IV) measurement to 0.1fA and 0.5µV
■Supports both quasi-static and medium-frequency capacitance-voltage (CV) measurement ]
■Supports accurate fast IV and time-domain measurement for a wide range of applications such as pulsed IV, NBTI and RTS measurement.
■Supports high voltage pulse generation (up to ±40 V) for high power and memory device testing.
Module Selection Guide:
| Module Main | Slots | Specification |
B1510A HPSMU | 2 | Up to 200V, 1A force. 10fA current resolution |
B1511A MPSMU | 1 | Up to 100V, 100mA force, 10fA current resolution |
B1517A HRSMU | 1 | Up to 100V, 100mA force, 1fA current resolution |
E5288A ASU | NA | Up to 100V, 100mA force, 100aA current resolutio |
B1520A MFCMU | 1 | 1kHz to 5MHz, up to 100V DC bias with SMU |
B1525A HV-SPGU | 1 | Min 12.5ns pulse width, 10ns transition time, up to 40V with 3 level pu |
B1530A WGFMU | 1 | Min 100ns pulse width, 10V peak-to-peak output, 5ns current or voltage measurement sampling speed |
Hot Tags: used semiconductor device analyzer agilen b1500a, China, manufacturers, Suppliers, Factory, buy, low price, in stock, Used Semiconductor Parameter Analyzer Agilent 4156B, Used Pulse Generator Unit Agilent B1525A, Used Mid Power Source Monitor Unit Agilent B1511A, Used DC Source Module Agilent 41420A, Used Pulse Generator Expansion Unit Agilent 41501A, Used Parametric Measurement System Mainframe Agilent E5270B